Purpose: To compare the results of Monte-Carlo simulations of proton radiography using energy resolved dose measurements with experiments.
Methods: A solid wedge of dimensions 35 cm x 35 cm (HxW) is kept in the path of a scanned pencil beam of protons. A flat panel detector kept behind the wedge, measures the dose. The energy of the beam is changed from 65 MeV to 250 MeV in steps of 5 MeV from which the energy resolved dose functions (ERDF) at different water equivalent thickness are estimated. This entire setup is simulated and the ERDF is retrieved in a similar manner. The simulation parameters like the thickness of the detector, noise background are changed to match the experimental data to the best extent possible.
Results: Preliminary results show a good match between experimental and simulation data provided optimal parameters are chosen in the simulation. The linearity of the dose response of the detector is a crucial factor that needs to be considered in the simulations.
Conclusion: We have studied here the comparison between the experimental and simulations of ERDF. From the preliminary results, we conclude that, it is indeed feasible to estimate the ERDF directly from simulations.