Purpose: To commission 10MV flattening-filter-free (10FFF) beam models with Eclipse for stereotactic radiosurgery (SRS) treatments.
Methods: AAA and AcurosXB 10FFF non-SRS customized models for a Varian Stx are used as baselines. The models are optimized in version 15.5 Eclipse (Varian, Palo Alto, CA). The focal spot, θ, and the minimum field size, FSmin, are (0.75mm,0.50mm) and 30x30mm² for both baselines. The θ, FSmin, output factors, St, and radial mean energy,
Results: The optimized θ of AAA and AcurosXB are (0.0mm,0.0mm) and (0.50mm,0.0mm) respectively. The FSmin and the corresponding St for both algorithms are 10.0x10.0mm² and 0.771. Tuning of
Conclusion: By carefully adjusting the input parameters, the accuracy of AAA and AcurosXB for small fields can be significantly improved for SRS treatments.