MENU

Click here to

×

Are you sure ?

Yes, do it No, cancel

Evaluation of a Novel EPID Based Dosimetry System for Verification of VMAT QA in Comparison with Traditional 3D Detector Quality Assurance

R Varadhan1*, S Way2, C Overbeck3, M Bedoya4, A Jahagirdar5, R Malmin6, R Malmin7, (1) Minneapolis Radiation Oncology, ,,(2) Minneapolis Radiation Oncology, ,,(3) Minneapolis Radiation Oncology, ,,(4) Minneapolis Radiation Oncology, ,,(5) Edina High School, ,,(6) Minneapolis Radiation Oncology,

Presentations

(Sunday, 7/12/2020)   [Eastern Time (GMT-4)]

Room: AAPM ePoster Library

Purpose:
EPID based dosimetry systems have been largely unavailable for patient dosimetry using the I-ViewGT detector in Elekta Linacs. The purpose of this study is to verify the accuracy of a novel EPID dosimetry system based on a convolution/superposition algorithm by benchmarking its performance against traditional 3D array detector system for VMAT QA.


Methods:
Ten patients with equal mix of 6MV and 10MV beam energies with varying degrees of VMAT plan modulation were randomly chosen for this study. The EPID dosimetry methodology generates a prediction model which reconstructs a 2D-dose map, at 5 cm depth in water at SAD level, from the DICOM RT Plan generated by the TPS. This is then compared to a conversion model which is used to compute a 2D dose map, related to 5cm depth in water at SAD level, from EPID signal values measured using the amorphous silicon detectors in I-ViewGT. The sensitivity of EPID dosimetry to known MLC errors was investigated by applying known MLC offset in both “in” and “out” directions for both leaf banks on the Agility 160 leaf MLC head. 3D detector QA was performed using the Delta4 system.

Results:
The gamma pass rates using the 3%/2mm global gamma criteria were greater than 95% for EPID dosimetry and greater than 92.5% for Delta4 measurements for all the selected VMAT plans.
When known MLC offsets were introduced by moving the MLC leaf banks “in” and “out” by 1mm each, the EPID dosimetry demonstrated adequate sensitivity to detect introduced errors.

Conclusion:
EPID based pretreatment quality assurance can be achieved with the EPIbeam system for fluence verification, and is comparable to traditional 3D detector based QA using the Delta4 system for routine VMAT QA.The system is sensitive to detect MLC errors on the order of 2mm.

Keywords

Quality Assurance, Electronic Portal Imaging, Intensity Modulation

Taxonomy

TH- External Beam- Photons: Quality Assurance - IMRT/VMAT

Contact Email