Submillimeter Monitoring of Intrafraction Patient Movement with Optical Surface Imaging

E Covington*, X Wu , I Brezovich , S Shen , R Cardan , R Popple , university Alabama Birmingham, Birmingham, AL

Presentations

(Sunday, 7/29/2018) 3:00 PM - 6:00 PM

Room: Exhibit Hall

Purpose: To evaluate the accuracy of monitoring intrafraction motion during stereotactic radiotherapy (SRT) with the Optical Surface Monitoring System (OSMS). Prior studies showing a false increase in the magnitude of translational offsets at non-coplanar table positions, which prompted the vendor to implement software changes. This study evaluated two software improvements intended to address false offsets.

Methods: The vendor implemented two software improvements: a volumetric rather than planar calibration (ACO) and, approximately 6 months later, an improved calibration workflow designed to better compensate for thermal drift (CIB). Offsets relative to the reference position obtained at table angle 0 following image-guided setup were recorded before beam-on at each table position and at the end of treatment when returned to table 0.

Results: Prior to ACO, between ACO and CIB, and after CIB, 223, 147, and 89 fractions were observed respectively. The median magnitude at the end of treatment was the same for all three intervals: 0.29 mm, 0.33 mm, and 0.31 mm. Prior to ACO, the offset magnitude for non-zero table positions had a median of 0.79 mm and was found to increase with increasing distance from isocenter to the anterior patient surface. After ACO, the median magnitude was 0.73 mm, but the dependence on isocenter location was eliminated. After CIB, the median magnitude for non-zero table positions was reduced to 0.52 mm.

Conclusion: Ongoing improvements in software and calibration procedures have decreased reporting of false offsets at non-zero table angles. However, the median magnitude for non-zero table angles is larger than that observed at the end of treatment, indicating that accuracy remains better when the table is not rotated.

Keywords

Optical Imaging, Stereotactic Radiosurgery